15.2
Warnings & Alarms
When an alarm is activated it will display on the screen and the vaporiser’s lights
will illuminate red. If the audible alarm is enabled this will sound, but only in
cycle.
The hazard symbol will be visible in the second box if there are any active alarms,
and will flash if there are unacknowledged alarms.
Touch box to view the alarm list. All alarms are printed if a printer is fitted.
The following is a table of all the alarms and what action can be taken:
Alarm
No.
Alarm Name
Alarm
Description
Action
User Action
1
Emergency Stop
The emergency stop
has been pressed
down
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Turn the Emergency
Stop button to reset
when safe to enter
room. See Section 4 for
location
2
RFID reader 1 error
Left hand RFID
reader alarm (reads
bottle labels)
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm and
contact service provider
3
RFID reader 2 error
Right hand RFID
reader alarm (reads
bottle labels)
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm and
contact service provider
4
Vaporiser 1
temperature low
Vaporizer not
reaching temperature
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
Alarm Acknowledge button. All alarms
have to be acknowledged to be removed
from the alarm list. To acknowledge first
touch the alarm warning in the panel
then press the acknowledge button
.
Acknowledging will stop the audible
alarm, if enabled.
To minimise or close an empty alarm list
touch the hazard symbol.