15
TROUBLE SHOOTING
15.1
Software Cycle Interlocks
The following conditions prevent the cycle from starting:
NOTE the ProteQ’s lights will be white, when the issue is resolved it will go green.
Insufficient H
2
O
2
Volume
If the bottle icons are yellow there is
insufficient volume of Hydrogen
Peroxide within the bottle to complete
the chosen cycle.
Replace a bottle or both bottles with
sufficient liquid for the cycle and repeat
Cycle Run procedure.
Invalid Hydrogen Peroxide
If the expiry date box is yellow and
flashing the hydrogen peroxide in the
bottle is beyond its expiry date and
cannot be used.
Replace bottle and repeat Cycle Run
procedure.
Bottle RFID tag not read or plunger not
down
If the bottle cannot be seen the cycle
will not run as the bottle RFID has not
been read. There is either no bottle in
place, the black plunger is not down or
the bottle is in place (with plunger
down) but the RFID tag is not in the
correct position.
If the tag is in the wrong place then the bottle is misaligned with the red stud.
Remove the bottle carefully and check that the bottom of the bottle has not been
punctured, and replace ensuring that the bottle is correctly aligned, see Section
7.2.