BIOPAC Systems BSLSTMA Product Sheet Download Page 1

 

 

 

PRODUCT SHEET 

[email protected] 

[email protected]

 

www.biopac.com 

 

 

BIOPAC Hardware   |   

STIMULATORS MP3X

   |   Page 1 - 9 

Updated: 9.12.2019

 

BSL STIMULATORS 

Modular Stimulators (0-100 V):

 

Low Voltage Stimulator/Adapter: 

 

BSLSTMB for MP36/36R/35 

OUT3 Output Adapter for built-in Stimulator (MP36 only)  

BSLSTMA for MP30/35/36/36R 

SS58L Low Voltage Stimulator (MP35 only)

 

 

See also

HSTM01

ELSTM1

ELSTM2

EL300S

 and 

EL400

 electrodes. 

 

BSLSTMB 

 

 

 

BSLSTMA 

 

 

 

Lab set up note

  

Placing the BSLSTMA/B unit too close to 
MP3X hardware can result in data distortion 
of the BSLSTMA/B pulse width signal; the 
distortion is more apparent at higher sampling 
rates. 

 

NEVER set the BSLSTMA/B atop an 

MP3X

 

 

Position the BSLSTMA/B away from the 

MP3X to reduce the signal distortion

 

Note 

The older “BSLSTM” uses dial reading and a flip range switch. The same guidelines and 
cautions described here apply, except when noted. 

The BSLSTM Stimulator works in conjunction with the Biopac Student Lab System to allow precise stimulus 
pulse outputting. Use the BSLSTM and the BSL 

PRO

 to perform a wide array of measurements, such as: 

 

Twitch sub-threshold & threshold 

 

Muscle tension/length vs. force 

 

Fatigue 

 

Maximum twitch responses 

 

Tetanic contraction 

 

Velocity 

 

Single twitch, summation 

 

Nerve conduction 

 

STIMULATOR PULSE DEFINITIONS 

 

Pulse width 

The time that the pulse is in the non-zero or active state.

 

Delay before first pulse

 

The initial delay from the start of acquisition to the start of the first pulse. 

Repetition period

 

The time between pulses, as measured from the start of one pulse to the start 
of the next pulse. This is the inverse of the Pulse rate.

 

Pulse rate 

The number of pulses that occur in a one-second interval, expressed in Hz. 
The 

Pulse rate

 relates to the 

Pulse period

 as follows: 

Summary of Contents for BSLSTMA

Page 1: ...ng and a flip range switch The same guidelines and cautions described here apply except when noted The BSLSTM Stimulator works in conjunction with the Biopac Student Lab System to allow precise stimul...

Page 2: ...menu in order to maintain direct Level recordings If the Range is changed during recording the user should manually enter a software marker to note the change by holding down F9 on a PC or Esc key on...

Page 3: ...cquisition unit The MP3X sends the Pulse width and Pulse rate information via this cable Manual Test button Used to diagnose problems with the BSLSTM stimulator unit When the Trigger and Reference Out...

Page 4: ...obtain the signal representing 0 V out of the stimulator 5 Add the Input value found in Cal 2 to the Input Value displayed for Cal 1 For example if Cal 2 is pressed and provides an Input Value of 255...

Page 5: ...r from the back of the BSLSTM into an open channel input port DB9 female connectors CH 1 CH 2 CH 3 or CH 4 on the front of the MP3X 7 Plug the 12 Volt DC adapter into the wall 8 Mate the DC output con...

Page 6: ...P W 100 Volts 400 ms 50 Volts 30 ms 1 ms P W 100 Volts 4 ms 50 Volts 3 ms Limits User adjustable lower and upper rate limits Resolution 2 microseconds Accuracy Better than 2 Initial Pulse Delay Time...

Page 7: ...Volts beyond functional limits 50 Volts 250 ms 10 ms P W 100 Volts 200 ms 50 Volts 150 ms 1 ms P W 100 Volts 20 ms 50 Volts 2 5 ms Limits User adjustable lower and upper rate limits Resolution 2 micr...

Page 8: ...lead with a BNC connector such as needle electrodes or clip leads for direct stimulation of animal or tissue preps Control the stimulus with the Output Control option of the BSL PRO software Output c...

Page 9: ...uld only be used under direct supervision of an Instructor Never place any stimulator leads in the mouth or any other body orifice To prevent a Ground loop the Ground of the stimulator electrode and t...