51
50
Cleaning and Disinfecting Procedures
Cleaning and Disinfecting Procedures
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with disinfecting wipes
listed above to clean any possible dirt, dust, blood and other body fluids.
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device thoroughly. (Note: All
blood and body fluids should be cleaned from surface before performing
disinfecting procedure)
3. Allow the surface to remain wet for 2 minutes.
3
NOTE
- Your
Rightest
GD500 Lancing Device has been tested to ensure that there is no change in the performance
of the device after 550 cleaning cycles and 550 disinfecting cycles. The testing simulates 2 cleaning and
disinfecting cycles per week over the typical life of the meter (5 years).
2
Allow to remain
wet for 2 minutes
1
CAUTION
- Users should wash their hands thoroughly with soap and water after handling the meter, lancing device or
test strips.
- Please examine your LCD screen, test strip port, buttons and surface of your meter and lancing device after
cleaning and disinfecting cycles. Stop using the meter and/or lancing device if any of the following occur:
- Thin, sliver streaks appear on the screen,
- The screen becomes cracked, soft, dissolved, brittle or swollen.
- You are unable to turn on/off your meter, operate the left/ right button, the lancing device
release button or depth adjustable cap.
- You are unable to enter meter settings, function modes or recall your testing results.
If you have any questions or concerns, please contact your
Rightest
Blood Glucose Monitoring System
GM232C authorized representative or call the Bionime customer service toll free at 888-481-8485.