The BM3 Vet Pro system is intended for use in the electromagnetic environment specified below.
The customer or the user of the BM3 Vet Pro system should assure that it is used in such an
environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic disc
harge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood, con
crete or ceramic tile. If floo
rs are covered with syntheti
c material, the relative humi
dity should be at least 30
%
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply line
s 1kV for input/output lin
es
2kV for power supply lin
es
1kV for input/output line
s
Mains power quality should
be that of a typical comme
rcial or hospital environmen
t.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality should
be that of a typical comme
rcial or hospital environmen
t.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic f
ields should be at levels ch
aracteristic of a typical locat
ion in a typical commercial
or hospital environment.
Summary of Contents for BM3Vet Pro
Page 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Page 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Page 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...