BM1VOM-1.24
3. SAVE RECORD 133
BM1VET Operation Manual
3.1 Outline
There are two modes to save data. One is called MONITOR mode. It saves the animal
’s ID/TYPE
without re-register once animal registered. The other called SPOT mode. It initializes the machine
once animal
’s record saved.
SPOT mode is good for measuring many animals. MONITOR mode is used for monitoring only one
animal constantly.
3.2 Adjust to Record Mode
Select setup icon in icon menu.
When SAVE MODE menu selected in menu widow, whenever touching touch wheel Key mode
switches to MANUAL and AUTO in turn.
MAKER
SERVICE
SYSTEM
KEY
SOUND:
ON
DISPLAY
SAVE
MODE:
MANUAL
USER
SERVICE
MAIN
MENU
3.3 Measure with Monitor Mode
Measure after setup mode to AUTO
MAKER
SERVICE
SYSTEM
KEY
SOUND:
ON
DISPLAY
SAVE
MODE:
AUTO
USER
SERVICE
MAIN
MENU
It saves measured data every 60 seconds.
Once NIBP measured, maintains measured data till the next measurement.
Not be able to delete measured Parameter data once save it in the machine completely Maintain ID
and TYPE after complete saving.
Alarm limit numeric value does not change after saving.
If additional NIBP measure did not occur in the next 60 seconds then it is regarded as NIBP
measurement was not performed.