Rev. 5.01
98
In general, the leads should be placed as clean as possible with the 60Hz noise minimized to
make it possible to generate a signal. To improve the RESP signal, use a 5-lead cable set (RL as a
neutral electrode). It is recommended that the lead be placed in the maximum expansion and
contraction range of the lung, especially if deep breathing is involved.
Avoid the liver and the ventricles of the heart to prevent 60Hz noise from pulsatile blood circulation
Note
Cables and connectors to measure respiration rate (RR) are commonly used
with ECG.
Display
①
Breathe indicator: indicates the detected breath.
②
Breathing number: displays the number of respirations per minute.
③
Respiration alarm limit: indicates respiration limits.
④
Apnea limit Setting: Apnea limit sign
Respiration settings
Menu
Description
Available settings
A. Respiration menu
A-1.
Alarm
RESP alarm setting menu
A-1-1.
Parameter alarm limit
RR, APNEA Parameter alarm, level,
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...