Rev. 5.01
95
COUPLET
1)
Occurs when two ventricular beats are detected and have non-ventricular
beats before and after the couplet. The coupling interval must be less than
600 milliseconds.
IRREGULAR
1)
Occurs when six consecutive normal R-to-R intervals vary by 100
milliseconds or more.
PAUSE
1)
Occurs when the interval between two consecutive beats exceeds three
seconds.
PVC
1)
Isolated premature ventricular complexes occur when a premature ventricular
beat is detected and has non-ventricular beats before and after.
R ON T
1)
Occurs when a ventricular complex is detected within the repolarization
period of a non-ventricular beat.
TRIGEMINY
1)
Occurs when two or more trigeminal cycles (a ventricular beat followed by
two non-ventricular beats) are detected.
V BRADY
1)
Ventricular bradycardia occurs when a run of three or more ventricular
beats is detected with an average heart rate that is less than or equal to 50
beats per minute.
1)
: BM5Vet Elite, BM7Vet Elite only
Note
The Brady limit matches the low heart rate limit. If the low heart rate limit is
changed, the Brady limit changes.
Arrhythmia Settings
Menu
Description
Available settings
A. ECG menu
A-1.
Arrhythmia
Arrhythmia parameter alarm setup
menu
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...