Rev. 5.01
87
ECG Cable color
AHA
:
American Heart Association (U.S.A.)
IEC
:
International Electrotechnical Commission (Europe)
3LEAD / 5LEAD
Lead wire
AHA
Color code
AHA
Label
IEC
Color code
IEC
Label
Right arm
White
RA
Red
R
Left arm
Black
LA
Yellow
L
Right leg
Green
RL
Black
N
Left leg
Red
LL
Green
F
V1(precordial)
Brown
V1
White
C1
ECG Signal Processing and Display
The monitor displays QRS Complex for an amplitude of 0.4 to 5.0 mV (0.2 -5.0 mV for scale
settings below 0.5 mV / cm) and a QRS width of 70 to 120 ms for large animals (also for small
animals with 40 to 100 ms). The heart rate is calculated from 15 to 300 times per minute using
the last 10 seconds of the R-R interval and the two longest intervals and the two shortest
intervals at the R-R interval. The remaining interval is averaged, and the current heart rate is
displayed in the HR parameter box of the main screen as a result.
If arrhythmia monitoring is possible, the HR parameter box will change accordingly. If you select
Basic, you can display three basic life treatening arrhythmias called ASYS, VFIB, and VTAC. If Full is
selected, a separate ARR parameter box will be displayed next to the HR parameter box (for
details on selecting the arrhythmia mode, refer to the Arrhythmia Setting chapter).
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...