Rev. 5.01
157
NIBP calibration
At least once every two years, or as specified
by local laws.
Mainstream and sidestream CO2
calibration check
At least once a year, or if you suspect the
measurement values are incorrect.
Battery maintenance
Battery
See thesection on Maintaining Batteries in
chapter 1.
Trouble shooting
Noise in ECG
Check that the filter settings are appropriate.
Check whether the electrode is attached well.
Check whether the gel on the electrode is dry.
The Patient’s skin Yes
is extremely dry
The electrode is Yes
in poor condition
The cable(lead wire) Yes
Is disconnected
Apply ECG gel or water to the area of
contact then use the electrode
Replace the electrodes (Use CE
compatable products
)
Replace lead wires
Repair the ECG Board
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...