User Manual Smart Storage Agitator 30
Version
: 1.0
Validation date of the manual: 01/08/2021
19 of 29
4.2
Hazards of RF radiation
The SST-A30 electronic system antennas each emit a frequency of 13.56 MHz with
a maximum power output of 4,05
dBμA/m at 3m (less than the 42dBμA/m limit
threshold).
ELECTROMEDICAL EQUIPMENT requires special precautions regarding EMC. The
SST-A30 must be installed and put into service according to the EMC information
provided by the ACCOMPANYING DOCUMENTS.
Portable or mobile RF communications devices may affect ELECTROMEDICAL
EQUIPMENT
The use of ACCESSORIES, transducers, and cables other than those specified, except
for transducers and cables sold by the EQUIPMENT MANUFACTURER or EM SYSTEM
as replacement parts for internal components, may result in an increase in
EMISSIONS or a decrease in the IMMUNITY of the DEVICE or EM SYSTEM.
The DEVICE or EM SYSTEM should not be used alongside other devices or stacked
with them.
The DEVICE or the EM SYSTEMS may be interfered with by other devices, even if
they comply with the CISPR EMISSION REQUIREMENTS.
4.3
Electromagnetic compatibility
The SST-A30 complies with applicable electromagnetic compatibility standards, however, the
user will ensure that any electromagnetic interference does not create an additional hazard,
such as radiofrequency transmitters or other electronic devices.
In this chapter you will find necessary information to ensure an installation and a
commissioning of the SST-A30 in best conditions in terms of electromagnetic compatibility.
The different cords of the SST-A30 must be distant from each other.
Some types of mobile telecommunication devices such as mobile phones are likely to interfere
with the SST-A30. The separation distances recommended in this chapter must therefore be
strictly observed.
The SST-A30 must not be used near or on another device. If this cannot be avoided, it must be
checked for proper operation under the conditions of use before use. Use of accessories other
than those specified or sold by Biolog-id as replacement parts may result in increased emission
or decreased immunity of the SST-A30.
The SST-A30 uses the 13.56 MHz frequency. The frequency band is 13.553 - 13.567 MHz in
accordance with the ISO 15693 standard. The modulation type is ASK and the RF mode is
TX/RX.
The maximum power output of the PRD_7150200A is 5.04
dBμA/m at 3m.
The tables below
are for the SST-A30 (PRD_7150200A).