11 of 29
2.4.3
RFID tag
The RFID tag is used to store product and patient data as well as PC bag traceability data.
Recommendation: For optimal localization, the tag should be centered on the bag.
Fig. 8: Platelet concentrate bag and RFID tag.
Fig. 9: Tag positioning.
You must place your RFID tag on the center of your bag so that the RFID tag cannot be within
2cm of the edges of the drawers of the SST-A30 (risk of non-detection of the RFID chip by the
antenna of the drawer).
The RFID tags compatible with the SST-A30 are passive tags. For more information, please
contact Biolog-id's Quality Department at
Tag
positioning
PC bag