GeneExplorer Thermal Cycler Operation Manual
Chapter Four Operation Instructions
-
18
-
the temperature ramp of reagent is lagging that of module, the actual temperature change of
reagent is inconsistent with our settings that the actual time of reagent reaching set temperature
is shorter than the set time (see the figure below for the relation between module temperature
and reagent temperature).
“Tube” mode considers the lagging reagent temperature ramp and provides an overshoot process
after reaching set temperature to enable reagent to reach the set temperature more quickly. If
setting time equal to that of “Block” mode, the actual time for reagent reaching set temperature
is obviously extended (see the figure below for the relation between module temperature and
reagent temperature).
4 File management
Click
Open Experiment
on the main interface to enter the interface of file management.
Click the dropdown list on the upper left to select disk for operation. For example, select USB
flash disk to operate on USB flash disk contents/GeneExplorer.
Press “
New Experiment
” to enter the interface of experiment parameter setting.
Press “
Run
” to enter the interface of run.
Press “
Export
” to export file or user.
Press “
Function
” to delete and rename file or user.
Module
Reagent
Module
Reagent