Chapter Two Characteristics
―
6
-
Chapter Two Characteristics
The chapter mainly introduces the operation, transport and storage conditions of the instrument,
and its basic parameters, performance and functions.
1
Normal operation conditions
Ambient temperature: 10
C ~ 30
C
RH: ≤70%
Power supply: 220V
~
50Hz
Altitude:
<
2km
Attention:
Before using the instrument, please confirm if the working conditions can
meet the above requirements. Pay special attention to reliable grounding of
the power line.
2
Transport and storage condition
Ambient temperature: -20
C ~ +55
C
RH: ≤80%
3
Basic parameters
The parameters of the modules are as below:
Name
Gene Explorer Thermal Cycler
Model
GE-96G
GE-384G
GE-48D
GE-4I GE-48DG
GE-48DS
GE-96S
Sample size
96×0.2ml
384×0.02ml
48×0.2ml+4
8×0.2ml
44×gla
ss slide
48×0.2ml+4
8×0.2ml
48×0.2ml+4
8×0.2ml
96×0.2ml
gradient
temperature
range
30℃~99℃
/
30℃~99℃
/
gradient
temperature
difference
range
1℃~30℃
/
1℃~30℃
/
Ladder
Temperature
range
/
0.1℃~6℃
Energy
consumption
600W
1200W
Fuse
250V 12A 5 20
LCD display
Color touch screen (8 inch, 260,000 colors)