Chapter 6 Parameter Monitoring
--94--
User Manual of Veterinary Monitor
Figure 6.50 AG Display
AG module can display all measured waveforms and parameters on the monitor screen, such as:
CO
2
: O
2
, N
2
O and AA1 waveform.
awRR: airway respiration rate.
MAC: minimum alveolar concentration.
CO
2
, O
2
, N
2
O and AA1: end-tidal (Et) and respiratory (Fi) values.
Where, AA1 represents one anesthetic gas of Des (desflurane), Iso (isoflurane), Enf (enflurane), Sev (sevoflurane)
and Hal (halothane).
AA1 waveform displays the main anesthetic gas waveforms.
Warning
Do not use flammable and explosive anesthetics such as ether and cyclopropane, so as to prevent
explosion.
6.10.3 MAC Value
Minimum alveolar concentration (MAC) is a standard for comparing the potency of inhalation anesthetics. 1
MAC represents the end-tidal concentration of an agent (at sea level) that, in 50 percent of a tested population,
prevents gross muscular movement in response to a painful, standardized stimulus.
The MAC value may be calculated and displayed by using end-tidal (Et) gas concentrations according to the
following formula:
Attention
Above formula does not consider the altitude.
6.10.4 Mainstream AG module
Intended Use
The IRMA main stream multi-gas probe is intended to be connected to other medical devices for display of
real time and derived monitoring data of CO
2
, N
2
O, and the anesthetic agents Halothane, Enflurane,
Isoflurane, Sevoflurane and Desflurane.
It is intended to be connected to an animal breathing circuit for monitoring of inspired/expired gases during
anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit and animal
room settings for
>
20 kg, 10
~
20 kg and
<
10 kg animals.
It is not intended to be used as the only means of monitoring an animal. It shall always be used in
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......