Appendix B Technical Specifications
-- 62 -- User Manual for Electrocardiograph
Using method A and D described in 4.2.7.1 of AAMI EC11 to test frequency response.
Because of sampling characteristics and the asynchronism between sample rate and
signal rate of the ECG machine, digital systems may produce a noticeable modulating
effect from one cycle to the next. This phenomenon, which is not physiologic, shall be
clearly described in the operator's and service manuals.
CMRR
>
115 dB
Animal Leak Current
<
10 μA
Sampling
rate
of
signals
8000Hz
B.1.2 Recorder Specification
Recorder
Thermal Dot Matrix Word Printing System
Recording Paper
50 mm, roll paper (iE 101)
80 mm, roll paper (iE 300)
Paper Speed
(5, 6.25, 10, 12.5, 25, 50)mm/s ± 5%
B.1.3 Other Specification
Display on LCD
5-inch TFT LCD screen (iE 101)
5-inch TFT LCD screen (iE 300)
Safety Classification
IEC60601-1, Class I, Type CF
AC Power Supply
100 V-240V, 50 Hz /60 Hz, 80 VA (iE 101)
100 V-240V, 50 Hz /60 Hz, 80 VA (iE 300)
DC Power Supply
Rechargeable lithium battery, 11.1 V/ 2600mAh.
In environment temperature 25
℃
± 5
℃
and with the machine turning off, the
charging time is not more than 2 hours to charge the battery to 90%.
In environment temperature 25
℃
± 5
℃
, the continuous working time is not less than
3 hours while the ECG device is continuously printing.
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......