2.1 Unpacking
inspection
II. Product installation
Check the packaging and instrument after unpacking the outer box of the Natch CS,
Please contact the transportation company or local distributor immediately if any
damage.
Check out the packing list for the accessories of the instrument main unit after
unpacking the outer box . Please contact the dealer immediately. whether the items are
complete if any inconsistency or missing.
2.2 Storage and transportation conditions and working environment
Storage and transportation conditions
Ambient temperature: -20
℃ ~
55
℃
.
Relative humidity: ≤93%.
Working environment
The installation environment should be as clean as possible without dust,
mechanical vibration, pollution, large noise sources, and electromagnetic
interference.
For indoor use, avoid direct sunlight or place it in front of heat and wind
sources. Choosea well-ventilated location.
Ambient temperature: 15
℃ ~
35
℃
.
Relative humidity: 15% to 90%.
Power requirements
Power supply: It must be used in a well-grounded power outlet. Before
installation,,Make sure the input voltage meets the requirements.
Voltage:
AC220V±22
V Frequency:
50Hz Power:
350VA
Space requirements
Leave enough space behind the machine and on both sides of the machine to prevent
poor heat dissipation which may cause the internal temperature of the machine to rise
and malfunction. The space where the instrument is placed must be larger than the
dimensions of the machineto ensure the space required to conduct maintenance, and turn
the power off in emergency.
The space must meet the following conditions, as shown in Figure 2-1:
Summary of Contents for BK-AutoHS96
Page 1: ...Automatic Nucleic Acid Extraction System BK AutoHS96 User Manual BIOBASE GROUP Version 2020 07...
Page 18: ...17 10 Rear service panel For mechanical arm maintenance...
Page 20: ...19 Figure 2 3 Internal Structure of Automatic Nucleic acid extraction system BK AutoHS96...
Page 23: ...22 7 Waste tip bag Collect abandoned tips...
Page 27: ...26 numbering function it is used to define the number of the first sample...
Page 52: ...51 Acid Extraction System or individual parts are...
Page 54: ...53 Appendix Parts List No Component name Specifications Quantity 1 Fuse 6 3A 2...