![BIO RAD ZE5 User Manual Download Page 53](http://html.mh-extra.com/html/bio-rad/ze5/ze5_user-manual_2753184053.webp)
Media Types
Legend
1
Stat tube position
Allows for loading a single tube, rather than a rack of
tubes, for quick sample acquisition without requiring
experiment configuration in the Experiment Builder.
2
Waste overflow port
(behind stat tube position)
Utilized for cleaning and wash station maintenance and
for removing clogs.
3
Clamp
Secures the tube rack or plate on the loader. Opens
when loader is extended. Closes when loader is
retracted.
4
Plate and tube rack position, agitation
assembly, and temperature control
Accommodates tube racks and plates. Integrates
agitation and temperature control.
5
Sample probe
Aspirates sample fluid and introduces it to the sample
pump for delivery to the flow cell.
6
Rinse line and rinse waste line
Introduces sheath fluid to the wash station for cleaning
the probe and sample line. During certain stages of
instrument operation such as shutdown, cleaning fluid,
rather than sheath fluid, can be used for washing.
7
Gauge
Supports the sample probe and the wash station.
8
Wash station
Facilitates probe and sample line cleaning.
9
Bead station
Houses beads used for quality control. Includes a cap
to prevent evaporation of the bead suspension fluid.
Integrated agitation ensures that beads are
resuspended before the QC process begins.
Media Types
The loader facilitates sampling from a variety of media types, including:
n
5 ml tubes (12 x 75 mm, 1–40 tubes per rack)
n
1.5 ml tubes (1–24 tubes per rack)
n
96-well plates
n
96 deep-well plates
n
384-well plates
n
One 5 ml tube in the stat tube position (12 x 75 mm)
User Guide | 51
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......