Chapter 2 Hardware Description
which results in higher side scatter (SSC). The ZE5 Cell Analyzer is configured to collect SSC from the
488 nm laser.
Optical Filters and Mirrors
The ZE5 Cell Analyzer’s optical filters separate and direct fluorescent light to the PMTs for detection.
The optical detection system is designed to be extraordinarily space-efficient and compact, allowing
up to 27 fluorescence detectors to be installed. A unique feature of the optical system design is that
each detector is an equal distance from the collection source, so that the system can collect every
wavelength of light at the same focal distance. A single lens and steering mirror are used to optimize
the collected light on all detectors, guaranteeing that the light efficiently reaches the final bandpass
filter with no more than three interactions with dichroic filters.
Mirrors and filters permit multiparametric analysis. By partitioning the spectrum of collected light into
specific ranges of wavelengths, each detector can be dedicated to the measurement of particular
fluorophores.
The ZE5 Cell Analyzer’s optical filters and detectors are arranged in banks dedicated to particular
interrogation point/laser combinations, with one bank per laser. These banks are located behind the
filter access door on the right side of the instrument. A black filter cover protects the optical detection
system from ambient light. When lifted out of the way, the filter cover attaches to the instrument using a
magnet, as shown in the next figure.
38 | ZE5 Cell Analyzer and Everest Software
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......