Fluidics System
To illuminate the sample area, press and hold the silver sample chamber button or click the Sample
Chamber Light button in the Instrument Tools dropdown located in the System section of the Everest
toolbar.
For more information about loading samples, see
.
Caution:
In addition to the loader door, there are two doors (filter access and bulk
fluidics access) that require access on a regular basis. Use caution when opening
and closing these doors and mechanisms to avoid pinching.
Flow Cell
The flow cell is the heart of the system and is where sample interrogation occurs. It is composed of
fused silica that surrounds a 145 x 265 µm channel through which sheath fluid flows and focuses the
sample fluid.
Sample is drawn up into the sample line by the sample pump and is introduced into the flow cell
through a sample introduction needle. Sheath fluid is introduced under pressure in an upward vertical
flow at 10 psi(~8 m/sec). The faster flow of sheath fluid around the sample hydrodynamically focuses
the sample into a narrow-diameter core stream. This allows cells to pass through at a high rate while
maintaining the integrity of single-file particle flow through up to five spatially separated laser
User Guide | 33
Summary of Contents for ZE5
Page 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Page 2: ......
Page 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Page 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Page 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Page 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Page 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Page 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Page 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Page 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Page 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Page 125: ...Working with Optical Filter Configurations User Guide 123 ...
Page 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Page 177: ...Setting Up the Run List User Guide 175 ...
Page 191: ...Setting Up the Run List User Guide 189 ...
Page 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Page 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Page 333: ...Decontaminating the System User Guide 331 ...
Page 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Page 397: ......