Other Guidelines
Swapping Fluidics During Sorting
: “Hot Swap”
New containers or bottles should be prepared before clicking the Swap Fluidics
button, as the swap time is limited to only 3 minutes before the system starting
to shut down due to low fluid levels.
Read carefully all procedure before starting the “
Hot Swap
”
1.
Click in the
Swap Fluidics
button of the
Setup and Maintenance
tab. The
Swap Fluidics
window will open