4.2.4 Error Messages
Error Message
Cause
Solution
ERROR STOP:
CODE 01
NO LOAD DETECTED
CORRECT AND CONTINUE
RUN PAUSED
CONTINUE CANCEL
1. Electrophoresis cell
not connected to the
power supply, or
buffer levels too low
2. Current load less
than 2 mA
1. Make sure all electrical
connections make good contact
and the cables and wire
electrodes are in good condition.
Verify buffer levels are
appropriate.
2. Verify the electrophoresis
application power requirements
match PowerPac Universal output
range.
NOTE:
The No Load detection
can be deactivated (see Sections
and
ERROR STOP:
CODE 02
OVER CURRENT
CORRECT AND CONTINUE
RUN PAUSED
CONTINUE CANCEL
Output leads shorted, or
there is shorting due to
wrong connections
Make sure all electrical connections
make good contact and the cables
and wire electrodes are in good
condition.
Verify buffer levels are appropriate.
ERROR STOP:
CODE 03
OVER VOLTAGE
CORRECT AND CONTINUE
RUN PAUSED
CONTINUE CANCEL
Control circuitry
problems
Press the CONTINUE soft key to
continue the run. If the problem
continues, cycle power and restart
the run. If the problem still persists,
contact Bio-Rad Technical Support.
ERROR STOP:
CODE 04
WATCHDOG TIMEOUT
Internal firmware failure Cycle power and restart the run. If the
problem still persists, contact Bio-Rad
Technical Support.
ERROR STOP:
CODE 05
STUCK KEY
Key on the front panel
stuck
Make sure keypad is not coated with
foreign substance. Try running
fingers over keypad to loosen keys. If
the problem persists, contact Bio-Rad
Technical Support.
ERROR STOP:
CODE 06
OUTPUT HIGH
1. Attempted restart
before output has
discharged
2. Very low loading
1. Wait longer before restarting.
2. Check connections for external
device driving the output.
If the problem persists, contact Bio-
Rad Technical Support.
ERROR STOP:
CODE 07
RUN POWER FAILURE
RUN TERMINATED
RESET WHEN READY
RESET
AC Power interrupted
during the run and the
run was terminated
because the Power
Failure detection is OFF
To continue a run after a power
failure, activate the Power Failure
Summary of Contents for PowerPac Universal Power Supply
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 12: ...Page 4 PowerPac Universal Power Supply Instruction Manual ...
Page 14: ...Page 6 PowerPac Universal Power Supply Instruction Manual ...
Page 30: ...Page 22 PowerPac Universal Power Supply Instruction Manual ...
Page 38: ...Page 30 PowerPac Universal Power Supply Instruction Manual ...
Page 40: ...Page 32 PowerPac Universal Power Supply Instruction Manual ...
Page 41: ......