BIO RAD Gel Doc XR+ Quick Start Manual Download Page 1

Gel Doc XR+ System

ChemiDoc XRS+ System

Fig. 1. Gel Doc XR+ and ChemiDoc XRS+ camera locations.

Fig. 2. Instrument Setup window.

1

  Installing Image Lab software version 5.1 beta

 After you have downloaded Image Lab software, proceed to 
installing the software by double clicking 

Setup

 

 Click 

Next

 to launch the installation wizard

Select 

“I accept the terms”

 and click 

Next

 

 Click 

Next

 to install Standard Edition 

Note:

  If you are in a regulated environment, you may wish to install the Security 

Edition. You will need to have purchased a Security Edition license from 
Bio-Rad Laboratories, Inc. to install the Security Edition. If you already 
have the Security Edition and have activated your license, reactivation of 
Image Lab is not necessary.

 Click 

Next

 to install to the default location or change to install  

to a different folder

 Click 

Install

 to begin installation. Click 

Finish

 to complete 

installation. Installation of Image Lab software version  
5.1 beta is now complete

 If you are upgrading a Gel Doc XR+ system, it is now stain-free enabled and ready to image stain-free 
gels. For upgrading Image Lab software on a ChemiDoc XRS+ system, a flat fielding calibration step 
will be required. Take a look at the camera on your imager if you are 
unsure which Bio-Rad imager you have (see Figure 1)

2

   Performing flat fielding calibration  

(optional step for Gel Doc XR+)

 To perform flat fielding you will need an orange fluorescent  
reference plate (catalog #170-8008) 

Note:

  To obtain an orange fluorescent reference plate, required for flat fielding 

calibration, contact Bio-Rad tech support at: 

[email protected]

 or 

1-800-4-BIORAD, option 2

.

Select 

Edit

 in the menu bar of Image Lab software

Then select 

Instrument Setup

 In the Flat Field section of the Instrument Setup window click 

Reset

  

(see Figure 2)

 Click 

Skip

 to bypass the lens flat fielding calibration because this 

procedure was done during installation of the imager

IMAGING

Stain-Free Software Upgrade

Quick Start Guide

Stain-Free Imaging on the Gel Doc

 XR+ and ChemiDoc

 XRS+ Systems

This guide will walk you through the steps to install Image Lab

 software version 5.1 beta.

Note:

  If you already have Image Lab software installed, then you will have to run the installer twice. First to uninstall the previous version, and then a 

second time to install version 5.1 beta. Alternatively, you may use Control Panel/Uninstall a Program to remove the previous version first.