Initial Preparation

1. 

Plan the plate layout.

2. 

Start up/warm up the Bio-Plex Multiplex Immunoassay System (

30 min

).

  Bring diluents, including wash buffer, assay buffer, standard  

diluent HB, detection antibody diluent HB, and sample diluent HB,  
to room temperature (RT). Keep the other items on ice until needed

 

– Mix by inversion to ensure all salts are in solution

 

– Prepare 1x wash buffer: dilute 

1 part

 10x wash buffer (60 ml) with  

9 parts

 distilled water (540 ml)

  Begin to thaw the frozen samples

3. 

 Prepare the sample dilution according to the guidelines provided in the 
following table. It is important to centrifuge serum or plasma samples at  

1,000 x g

 for 

15 min

 at 

4°C

 to remove particulates from all samples prior to use.  

Sample Type

Recommended Sample Dilution

Diluent

Serum and plasma

1:4

Sample diluent

Culture media and fluids

User defined

D 0.5% bovine 
serum albumin (BSA) (w/v)

Note:

 ICAM-1 and VCAM-1 require higher dilution for serum and plasma (recommended 100-fold). Refer 

to the Bio-Plex Pro Human Immunotherapy Assays Instruction Manual (#10000112563) for detailed sample 
preparation recommendations.

4. 

Calibrate the Bio-Plex System within Bio-Plex Manager Software. 

5. 

Reconstitute the standards and control by adding 250 µl of standard 
diluent HB to each. 

Vortex

 at medium speed for 

5 sec

 and incubate all 

vials on ice for precisely 

30 min

Bio-Plex Pro

 

 Human Immunotherapy Panel

Quick Guide

  For Use with 

Instruction Manual  # 

  Bio-Plex Pro Human Immunotherapy Assays 

10000112563

This guide can be used to prepare and run a full 1 x 96-well assay plate. New 
users can download the complete manual, which includes detailed instructions 
and a list of kit components, a

bio-rad.com/bio-plex

.