UF V (E3) 09/2018
Page 85/116
22. Data monitoring and recording
22.1 Ethernet interface
The chamber is regularly equipped with an Ethernet interface (8) that can connect the BINDER communi-
cation software APT-COM™ 3 DataControlSystem (option). The MAC Address is indicated in the “Ether-
net” controller menu (chap. 15.1.1).
22.2 Communication software APT-COM™ 3 DataControlSystem (option)
You can connect the BINDER communication software APT-COM™ 3 DataControlSystem via the cham-
ber’s Ethernet interface (8).
The APT-COM™ System permits the networking of up to 40 chambers and connection to a PC for con-
trolling and programming, as well as recording and representing temperature data.. The actual tempera-
ture value is given at adjustable intervals. Programming can be performed graphically via PC. For further
information on networking please refer to the operating manual of the BINDER communication software
APT-COM™ 3.
22.3 Analog output for temperature (option)
With this option the freezer is equipped with an analog output 4-20 mA for temperature. This output per-
mits transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (10) in the rear connection panel as follows:
ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Tempe
Temperature range: +40 °C /
104°F
to -100 °C /
-148°F
A suitable DIN plug is enclosed.
Figure 24: Pin allocation of DIN socket (10) for option analog output
22.4 Data logger kit (option)
The BINDER Data Logger Kit offers an independent long-term measuring system for temperature. It is
equipped with a keyboard and a large LCD display, alarm functions and a real-time function. Measure-
ment data are recorded in the Data Logger and can be read out after the measurement via the RS232
interface of the Data Logger. It offers a programmable measuring interval and permits storing up to 64000
measuring values. Reading out is done with the Data Logger evaluation software. You can give out a
combined alarm and status protocol directly to a serial printer.
Data Logger Kit T 220:
Temperature range -90 °C /
-130 °F
up to +220 °C /
428 °F
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the
manufacturer, supplied with the data logger.
Summary of Contents for ULTRA.Guard UF V 500
Page 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Page 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Page 109: ...UF V E3 09 2018 Page 109 116...
Page 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Page 117: ......