MK / MKT (E5) 03/2019
page 104/158
18. Anti-condensation protection via operation line
The anti-condensation protection condensates the chamber humidity at the coldest point in order to avoid
the samples becoming wet from condensation. Anti-condensation protection is performed by the evapora-
tor and can be programmed On/Off via operation line “Anti-condensation” in Fixed value and program
modes.
Use the anti-condensation protection only if absolutely necessary to prevent condensation on
the charging material.
When the anti-condensation protection is enabled the refrigeration machine keeps operating within warm-
ing-up phases (On = refrigeration machine operating, Off = refrigeration machine off).
•
If possible, use the anti-condensation protection only during warm-up phases.
If necessary it
can also be activated during hold phases.
•
Do NOT use the anti-condensation protection above a temperature set-point of +20 °C / 68 °F
maximum.
To obtain optimal warming results without condensation on the samples, program a heating gradient of
approx. 0.5 °C/min.
Example:
W/°C
t/min.
00
01
02
03
-40
-20
0
20
40
60
80
120
180
380
04
ON
OFF
t/min.
Minimum
tolerance
Maximum
tolerance
Operation line “Anti-condensation“
Depending on size, material, and shape of the charging material and on the heating-up rate, condensation
may form despite the activated anti-condensation protection. This condensation is, however, reduced
compared to the state without anti-condensation protection.
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......