KBF-S (E6) 08/2018
page 60/88
20. Options
20.1 Communication software APT-COM™ 3 DataControlSystem (option)
The chamber is regularly equipped with an Ethernet interface (6) that can connect the BINDER commu-
nication software APT-COM™ 3 DataControlSystem. The actual temperature and humidity values are
given at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers can
be cross-linked. The MAC Address is indicated in the “Ethernet” controller menu (chap. 14.1.1). For fur-
ther information on networking please refer to the operating manual of the BINDER communication soft-
ware APT-COM™ 3.
20.2 Data logger kits (option)
BINDER Data Logger Kits offer an independent long-term measuring system for temperature and humidi-
ty, available for different temperature ranges. According to the selected kit, the Data Logger can measure
and record also the ambient temperature and humidity values via a second multi-function sensor.
BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a
real-time function. Measurement data are recorded in the Data Logger and can be read out after the
measurement via the RS232 interface of the Data Logger. It offers a programmable measuring interval
and permits storing up to 64000 measuring values. Reading out is done with the Data Logger evaluation
software. You can give out a combined alarm and status protocol directly to a serial printer.
Data Logger Kit TH 70:
Multi-function sensor for chamber temperature and humidity: Temperature range
-40 °C /
-40 °F
up to 70 °C /
158 °F
, humidity range 0% r.h. up to 100% r.h.
Data Logger Kit TH 70/70:
Multi-function sensor for chamber temperature and humidity: Temperature
range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity range 0% r.h. up to 100% r.h. Multi-function sensor
for ambient temperature and humidity: Temperature range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity
range 0% r.h. up to 100% r.h.
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of
the manufacturer, supplied with the data logger.
20.3 Water protected internal socket (option)
The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Heat emission of electrical devices connected inside the chamber may modify the temperature
and humidity range.
CAUTION
Risk of short circuit.
Damage to the chamber.
Use the supplied plug only (IP protection type 67). Plug it in and tighten it by screwing
to secure contact.
If the socket is not used, close the screw lid and turn it to secure.
Summary of Contents for KBF-S 240
Page 75: ...KBF S E6 08 2018 page 75 88 24 7 Dimensions Dimensions size 240...
Page 76: ...KBF S E6 08 2018 page 76 88 Dimensions size 720...
Page 78: ...KBF S E6 08 2018 page 78 88...
Page 79: ...KBF S E6 08 2018 page 79 88...
Page 81: ...KBF S E6 08 2018 page 81 88...
Page 82: ...KBF S E6 08 2018 page 82 88 26 Product registration...
Page 89: ......