KBF / KBF P (E5.2) 02/2012
page 26/103
4.2 Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (18) on the rear of the unit (olive
∅
14
mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it has
to be permanently resistant against at least 95 °C /
203 °F
.
•
Mount the wastewater hose with a maximum positive inclination of 1 m /
3.28 ft
and a maximum total
length of 3 m /
9.8 ft
.
•
Protect both ends of the drainage hose with two of the four supplied hose clamps.
Wastewater is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous wastewater flow.
(15)
(18) (17)
(15) Power cable
(16) not used
(17) Freshwater connection “IN” with screw
thread ¾’’ for hose ½“, with union nut
(18) Wastewater connection “OUT” with hose
olive for hose ½“
Figure 11: Rear view KBF / KBF P with water connections
Protect the wastewater supply at both sides with the supplied hose clamps.