FED (E2) 12/2014
page 20/55
5.
Press
button to return to normal display (actual value display) (automatically after approx. 30
seconds).
Adjust the temperature safety device following any changes of the set-point (chap. 7).
6.2 Display / entry of temperature and ventilation set-points (with selected
temperature ramp)
If previously a temperature ramp value has been selected (chap. 6.4.2):
Press button
in normal display / actual value display during ramp operation to have displayed the
actual temperature ramp mset-point changing according to the selected gradient in addition to the
entered final set-points for temperature and fan speed.
The oven is operating, the controller is in normal display (actual value
display). The
actual temperature value
(example: 22 °C) is displayed:
1.
Press
button
The display shows alternately “
SPr
” and the
actual temperature ramp set-point
changing
according to the selected gradient (example: 42 °C):
This ramp set-point is only displayed, not adjustable.
2.
Press
button
The display shows alternately “
SP
” and the previous
temperature set-point
(example: 60 °C):
3.
With the
buttons enter a set-point value between 0 and 300.
The desired temperature set-point can be selected in a temperature range from 5 °C
above room temperature up to 300 °C.
Wait 2
seconds until the entered temperature value
is
taken over
(display flashing once).
4.
Press
button to proceed to the fan speed entry.
The display shows alternately “
n
” and the previous
fan speed set-point
(example: 100%):