9.
Time programs
The MB2 program controller permits programming time programs with real-time reference. It offers 25
program memory positions with up to 100 program sections each.
For each program section you can enter set-points for temperature, CO
2
, O
2
(chamber with O
2
control),
and humidity (chamber with active humidification), section duration, type of set-point transition (ramp or
step) and the tolerance ranges.
If the safety controller has been set to “limit” mode, check the setting of the safety controller
when changing the temperature set-point, (chap. 12.2).
Programming remains saved in case of a power failure or after turning off the unit.
Path:
Main menu > Programs> Time program
9.1
Starting an existing time program
In Normal display press the
Program start
icon to access the “Program start” menu.
“Program start” menu
•
In the field “Program type” select the setting “Time program”.
•
In the field “Program” select the desired program.
•
Select the field “Program start” and enter the desired program start time in the “Program start” entry
menu. Press the
Confirm
icon. The program delay time until program start begins.
The program end is adapted automatically depending on the entered program duration.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu. The
program starts running.
If instead you press the
Close
icon to exit the menu without taking over the entries, the program will not
start.
Normal display. Information on the bottom of the
screen indicates the currently running program and the
time already passed. The grey bar shows how much
time of the whole time is elapsed. If program duration
has been set to infinite, the grey bar is not displayed.
Summary of Contents for CB 170