16.2 Setting the parameters
This menu allows setting the storage interval, the type of values to be shown and the scaling of the
temperature and humidity charts.
Path:
Main menu > Settings > Measurement chart
“Measurement chart” submenu.
•
Select the field “Storage interval” and enter the desired storage interval. Confirm entry with
Confirm
icon.
The available presentation depends on the pre-selected storage rate. Factory setting: 60 seconds.
This means the higher the storage rate, the more precisely but shorter the data representation will be.
•
In the field “Storage values” select the desired
value type to be displayed.
•
For scaling the representation select the desired minimum and maximum temperature or humidity
value and enter the desired values. Confirm each entry with
Confirm
icon.
Display ranges:
•
Temperature: 0 °C up to 200 °C
•
CO
2
: 0 Vol.-% up to 100% Vol.-%.
•
O
2
: 0 Vol.-% up to 100% Vol.-%.
•
Humidity: 0% r.h. up to 100% r.h.
CAUTION
Setting the storage rate or rescaling (minimum and/or maximum) will clear the
measured-value memory and the event list.
Danger of information loss.
Change the storage rate or scaling ONLY if the previously registered data is no longer
needed.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.