9. Ramp function
9.1 General information
You can program temperature ramps in order to extend heating up times. This may be necessary in some
cases to prevent temperature stress in the material during the heating up phase. Temperature ramps should
only be used if required. Using them may result in considerably slowing down the heating up times. When
the ramp function is turned off, the chamber will heat up with its maximum heating capacity.
The entry means the nominal value gradient and limits the maximum temperature increase to this value.
Due to the heat and evaporation energy assumed by the drying material, smaller temperature gradients
may also result.
A ramp proceeds from a previously entered set-point to a new, higher one. The temperature must be equil-
ibrated to the start set-point. Perform the setting in the following 3 steps:
1.
Enter the temperature set-point as
start ramp set-point
and let the temperature equilibrate to this value
2.
Define the temperature increase (
ramp gradient
) in °C/min or in °F/min in the setting menu “Ramp
function”
You can select a gradient from „0.0“ up to „1.0“ or from „1“ up to „10“ according to the chamber type.
When setting the gradient to „0.0“ or „0“, ramp function is turned off. The chamber will then heat up with
its maximum heating capacity.
The chamber will try to heat up according to the entered gradient, i.e. with a speed of xx degree per
minute. A heating-up rate of 0.4 °C/min for the incubators BD and BF resp. 4 °C/min for the heating and
drying ovens ED, FD and FED can be regarded as a realistic maximum.
3.
Enter the
target ramp set-point
in the “ramp function” setting menu.
As soon as the entries have been adopted, the ramp function is activated. The chamber heats up with the
entered gradient, if the set ramp target value lies above the actual temperature value.
During ramp operation the
effective ramp set-point
continually rises in accordance to the entered gradient
from the previously entered set-point to the new one. The actual value follows this continually changing
effective ramp set-point. As soon as the ramp target value is reached, this temperature is maintained con-
stant.
The actual temperature value, the selected gradient, and the target value are shown in Normal display.
The effective ramp set-point can be seen through the temperature set-point function.
Summary of Contents for BD 115
Page 86: ...BD BF ED FD FED E3 1 02 2022 page 86 116 16 11 Dimensions size 56...
Page 87: ...BD BF ED FD FED E3 1 02 2022 page 87 116...
Page 88: ...BD BF ED FD FED E3 1 02 2022 page 88 116 16 12 Dimensions size 115...
Page 89: ...BD BF ED FD FED E3 1 02 2022 page 89 116...
Page 90: ...BD BF ED FD FED E3 1 02 2022 page 90 116 16 13 Dimensions size 260...
Page 91: ...BD BF ED FD FED E3 1 02 2022 page 91 116...
Page 92: ...BD BF ED FD FED E3 1 02 2022 page 92 116 16 14 Dimensions size 720...
Page 93: ...BD BF ED FD FED E3 1 02 2022 page 93 116...
Page 94: ...BD BF ED FD FED E3 1 02 2022 page 94 116...
Page 96: ...BD BF ED FD FED E3 1 02 2022 page 96 116...
Page 97: ...BD BF ED FD FED E3 1 02 2022 page 97 116 17 2 EU Declaration of Conformity for BF...
Page 98: ...BD BF ED FD FED E3 1 02 2022 page 98 116...
Page 99: ...BD BF ED FD FED E3 1 02 2022 page 99 116 17 3 EU Declaration of Conformity for ED...
Page 100: ...BD BF ED FD FED E3 1 02 2022 page 100 116...
Page 101: ...BD BF ED FD FED E3 1 02 2022 page 101 116 17 4 EU Declaration of Conformity for FD...
Page 102: ...BD BF ED FD FED E3 1 02 2022 page 102 116...
Page 103: ...BD BF ED FD FED E3 1 02 2022 page 103 116 17 5 EU Declaration of Conformity for FED...
Page 104: ...BD BF ED FD FED E3 1 02 2022 page 104 116...
Page 106: ...BD BF ED FD FED E3 1 02 2022 page 106 116...
Page 107: ...BD BF ED FD FED E3 1 02 2022 page 107 116...
Page 108: ...BD BF ED FD FED E3 1 02 2022 page 108 116...
Page 109: ...BD BF ED FD FED E3 1 02 2022 page 109 116...
Page 117: ......