KBF-LQC (E5.2) 12/2010
page 86/106
17.10 BINDER Pure Aqua Service (option)
The optional BINDER water treatment system (disposable system) serves to treat tap water. The lifetime
depends on water quality and the amount of treated water. The measuring equipment to assess the water
quality is reusable.
For detailed information on operating the water treatment system BINDER Pure Aqua Ser-
vice and its function, please refer to the operating manual Art. No. 7001-0159, delivered
with BINDER Pure Aqua Service.
18. Maintenance, cleaning, and service
18.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅
The unit must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the unit.
¾
Disconnect the unit before conducting maintenance work. Disconnect the power plug.
¾
Ensure all maintenance work is conducted by licensed electricians or experts author-
ized by BINDER.
Ensure regular maintenance work is performed at least once a year.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Have conducted regular maintenance work on the steam humidifier at least once a year. The operating
behavior and the maintenance intervals of the humidifier essentially depend on the available water quality
and the amount of steam produced in the meantime.
We recommend cleaning the condensers at least twice a year. A qualified technician must
perform cleaning.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing)
several times a year.
BINDER factory service provides worldwide calibration, and adjustment services for both light sensors
without need for expensive optical measuring devices at the service points. Adjustment can be done with
a portable and adjustable source of current. A service technician selects the current values at the meas-
uring channel which the sensor reaches at defined light values. Then a measured value correction is
executed at the MB1 controller equivalent to that of temperature and humidity, according to the calibration
certificates of the sensors.