25. Zero-voltage switching outputs via operation lines
The chambers are equipped optionally with four zero-voltage switching outputs (DIN sockets (17) and (18)
located in the lateral control panel).
The operation lines serve to switch any device connected to the zero-voltage relay output. They can be
programmed On/Off in Fixed value and program modes.
Connection for operation lines “Switching output 1” and “2” occurs via DIN socket (17), connection for op-
eration lines “Switching output 3” and “4” via DIN socket (18) in the lateral control panel:
OUTPUT TRACK 1+2
24V/MAX.2,5A
OUTPUT TRACK 3+4
24V/MAX.2,5A
Figure 24: Pin configuration of DIN sockets (17) left and (18) right
DIN socket (17):
Operation line “Switching output 1”
Operation line “Switching output 2”
Pin 1: Pin
Pin 2: Make
Pin 4: Pin
Pin 5: Make
DIN socket (18):
Operation line “Switching output 3”
Operation line “Switching output 4”
Pin 1: Pin
Pin 2: Make
Pin 4: Pin
Pin 5: Make
Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5 A
DANGER
Electrical hazard through overload of contacts.
Deadly electric shock. Damage to the switching contacts and connection socket.
∅
Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.
∅
Do NOT connect any devices with a higher loading capacity.
1
2
4
5
1
2
4
5
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Page 173: ......