MK LIT (E5) 01/2021
page 135/172
22.2 Setting the parameters
This menu allows setting the storage interval, the type of values to be shown and the scaling of the tem-
perature charts.
Path:
Main menu > Settings > Measurement chart
“Measurement chart” submenu.
•
Select the field “Storage interval” and enter the desired storage interval. Confirm entry with
Confirm
icon.
The available presentation depends on the pre-selected storage rate. Factory setting: 60 seconds. This
means the higher the storage rate, the more precisely but shorter the data representation will be.
•
In the field “Storage values” select the desired
value type to be displayed.
•
For scaling the representation select the desired minimum and maximum temperature value and enter
the desired value. Temperature display range: -50 °C (MK) up to 110 °C. Confirm entry with
Confirm
icon.
Setting the storage rate or rescaling (minimum and/or maximum) will clear the measured-value memory
and the event list.
NOTICE
Danger of information loss when setting the storage rate or rescaling.
Data loss of measured-value memory and event list.
Change the storage rate or scaling ONLY if the previously registered data is no longer
needed.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Page 173: ......