MK LIT (E5) 01/2021
page 125/172
20.5.2.2 Display of MAC address
Path:
Main menu > Device info > Ethernet
“Ethernet” submenu (example).
20.5.3 Web server
This controller menu serves to configure the web server. Then you can enter the chamber’s IP-address in
the Internet. The IP address is available via
Chamber information > Ethernet
. The BINDER web server
opens. Enter the user name and password which have been assigned for the web server in the controller
menu. This enables online access to the controller display, to see e.g., the event list or error messages. In
this view no settings can be changed.
Path:
Main menu > Settings > Web server
“Webserver” submenu.
•
In the field “Password active” select the desired
setting “Yes” or “No”.
•
Select the field “User name” and enter the desired user name. Confirm entry with
Confirm
icon.
•
Select the field “Password” and enter the desired password. Confirm entry with the
Confirm
icon.
•
Select the field “Automatic log out after” and enter the time in minutes after which the webserver shall
log out automatically. Setting range: 0 min to 65535 min. Confirm entry with
Confirm
icon.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...