14.4 Maintenance work by the customer
14.4.1 Checking and cleaning the condenser air filter
The condenser air filter prevents accumulation of dust on the condenser. If the filter is blocked by dust
this may cause decrease or failure of refrigeration.
Check the air filter visually for soiling every month. Especially with the alarm message “A52” (chap. 17.2)
the filter may be soiled. You can rinse the filter and use it again.
Regularly check the filter visually for soiling.
The filter is located behind a grille (D) in the lower housing cover. You can easily take it out for cleaning.
Open the grille by pushing the provided recess.
Figure 69: Opening the filter grille
Pull out the filter upward straightly out of the grille.
Figure 70: Taking out the filter
Wash the filter with water and let it dry.
Fix the filter correctly following cleaning.