8.2
Operating mode HAND
•
In Normal display, press button “MODE” to toggle to operating mode HAND.
•
Use the “MODE” button to successively call up the operating functions.
•
Use “▲”
or “▼” to set the values of the operating functions.
Operating functions that can be set in operating mode HAND
Display Setting range
Operating function
-90 °C to -40 °C
Temperature set point
-70 °C to -20 °C
Set-point entry of CO
2
emergency cooling
(only with option CO
2
emergency cooling “On”)
-86 °C to -40 °C
Safety controller set point when set point type is “Limit” (Lit)
Limit value, i.e., maximum permitted absolute temperature value. When
exceeded, the safety controller triggers an alarm. Set the limit value by
approx. 15 K above the controller temperature set point.
Indication of supply voltage
(only with option Advanced voltage boost-
er)
0 to 999
Max. 3-digit numeric
password
for locking/unlocking the operating func-
tion setting in operating mode LOCK (chap. 8.4)
To permit temporary unlocking: select any password other than 0.
To permit permanent unlocking: set password to “0”.
Set the safety controller each time the set point for the temperature is changed. Set the safety
controller set point by approx. 15 K above the controller temperature set point.
Summary of Contents for 9020-0213
Page 6: ...UF V E2 E2 1 03 2013 page 6 112...
Page 7: ...UF V E2 E2 1 03 2013 page 7 112 Product registration...
Page 104: ...UF V E2 E2 1 03 2013 page 104 112 18 7 Dimensions UF V 500 E2 xxx Indications in mm...
Page 105: ...UF V E2 E2 1 03 2013 page 105 112 18 8 Dimensions UF V 500 E2 1 xxx Indications in mm...
Page 106: ...UF V E2 E2 1 03 2013 page 106 112 18 9 Dimensions UF V 700 E2 1 xxx Indications in mm...
Page 113: ......