MK (E3) 07/2009
page 51/70
Instrument panel
Standard commercial cleaning detergents free from acid or halogenides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
Decontamination
Disconnect the chamber from the power supply prior to decontamination. Disconnect the power plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or halogenides.
Alcoholic solutions.
We recommend using disinfectant Art. No. 1002-0022.
In case of contamination of the interior by biologically or chemically hazardous goods, there are two pos-
sible procedures depending on the type of contamination and charging material.
(1)
Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during
the decontamination process.
(2)
If necessary, have strongly contaminated inner chamber parts removed by an engineer for clean-
ing, or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
With every decontamination method, always ensure adequate personal safety.
CAUTION
Danger of corrosion.
Damage to the unit.
∅
Do NOT use acidic or chlorine cleaning detergents.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
15.3 Sending back the unit to BINDER GmbH
If you send a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. We will issue an authori-
zation number after receiving your complaint either in writing or by telephone prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
mentioned below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of defect or fault
•
Your full address; if possible contact person and availability of that person