Repair work at the outer door
6.2
Checking and replacing outer door components
6.2.1
Dismounting the interior side of the outer door
Necessary tools: Crosstip screwdriver, size 2
1. Disconnect the unit from the power supply.
2. Remove the door gasket (chap. 6.2.6,).
3. Remove the screws (F) of the holders (E) of the interior side of the door.
4. Lift the holders (E) over the fastening spigot (G), so that you can remove the holders.
5. Slightly incline the interior side of the door (C) on the top at the frame and pull the plug of the
grounding cable (D) from the connection on the top right-hand at the interior side of the door.
6. Lift the interior side of the door (C) to lift its fastening spigots at the bottom side out of the bore
holes in the door frame (A).
7. The interior side of the door (C) is still connected to the unit by the cables (B) of the door heat
conductor and of the door temperature sensor.
8. Carefully place the interior side of the door (C) on the floor in front of the door frame (A) and
lean it against the unit.
9. Mount the interior side of the outer door by following steps 1 to 8 in reverse order
CAUTION!
Danger of damage.
Electric cables are not made for holding larger components and are damaged by doing this.
∅
Do not let the interior side of the outer door (C) hang at the cables of the door heat conductor or of the
door temperature sensors.
Service Manual KBF (E5.2) 04-2015
43