Decontamination
Disconnect the chamber from the power supply prior to decontamination. Pull the power plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or halogenides.
Alcohol based solutions.
We recommend using disinfectant Art. No. 1002-0022.
In case of contamination of the interior by biologically or chemically hazardous goods, there are three
possible procedures depending on the type of contamination and charging material.
(1) The safety / paint drying ovens FDL can be hot air sterilized at 190 °C /
374°F
for at least 30 minutes.
All inflammable goods must be removed from the interior before.
(2) Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during
the decontamination process.
(3) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning,
or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
With every decontamination method, always ensure adequate personal safety.
CAUTION
Danger of corrosion.
Damage to the unit.
∅
Do NOT use acidic or chlorine cleaning detergents.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
14.4 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone
prior
to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of the defect or fault
•
Complete address, contact person and availability of that person
•
Exact location of the BINDER product
in your facility
•
A contamination clearance certificate (chap. 18) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recog-
nized or be recorded clearly in the delivery documents.
For security reasons we cannot accept a unit delivery if it does not carry an authorization num-
ber.