Display 1
Display 2
Figure 7: RD3 program controller
LED function indications and their signification:
(3a) (yellow)
Heating
active
(3b)
(yellow)
No function
(3c)
(yellow)
No function
(3d)
(green)
Illuminates: program operation
Flashes: exceeding of the tolerance limits in “Fixed value entry mode” or in
“Program operation”. In “Program operation”: program interruption.
The program controller RD3 permits programming of temperature cycles.
Two programs with up to 10 sections each or one program with up to 20 sections can be entered (setting
in the user level, chap. 10).
When changing from 2 programs to 1 program or vice-versa, existing programs are deleted
The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs 59
min (setting in the user level, chap. 10). This setting is then valid for all program sections.
Programming can be done directly via the controller keyboard or graphically at the computer using the
software APT-COM™ 3 DataControlSystem (option, chap. 13.1) specially developed by BINDER.
5.2
General indications
The program controller RD3 offers several functional levels:
Normal Display / fixed value operation:
•
Display of the actual value of temperature (display 1) and of the actual date and time (display 2).
•
The chamber is in fixed value operating mode, adjusting to the entered set-points.
Fixed value entry mode
•
Entry of the temperature set-point for fixed value operating mode
•
Entry of temperature set-points 1 and 2 for week program operation