FED / FED-UL (E2) 01/2022
page 28/60
1.
Press down
button for approx. 5 seconds.
The display alternately shows “unit” and the temperature unit:
2.
Press again button
.
The display alternately shows “
rASd
” and the actual setting of the
set-point gradient
:
3.
Set the desired ramp gradient with buttons
(set-point gradient in °F or °C acc. to setting in
chap. 6.4.1).
The set value is automatically adopted after 2 seconds.
During ramp operation the actual set-point (SPr) continually rises in accordance to the entered gradient
from the previously entered set-point to the new one (SP). The actual value follows the set-point value.
About set-point display during ramp operation see chap. 6.2.
6.4.3 Chamber addressing
If several chambers are networked with a PC via the APT-COM™ 4 Multi Management Software (option,
chap. 8.1), each chamber must be allocated a unique address. Addressing takes place on the chamber
controller as follows:
1.
Press down
button for approx. 5 seconds.
The display alternately shows “unit” and the temperature unit:
2.
Press again button
.
The display alternately shows “rASd” and the set-point gradient:
3.
Press again button
.
The display alternately shows “
Adr
” and the actual setting of the
chamber address
: