WARNING
To not route the terminal connection cables alongside power
cables as these could cause interference that could prevent
the terminal from operating in the correct way. Only use the
connection cable supplied with the terminal. If the cable
supplied is too short, do not attach an extension lead but
contact the Manufacturer.
3.2 Connection of the terminal to the weighing platform
To connect the terminal to the weighing platform, the supply normally
includes a pre-wired cable whose female connector must be inserted
into the 9-pin male terminal (JBIL1 or 2) at the rear of the indicator (see
point 5 Figure 3.2 on page 1-20). In the case of a duplex scale, scale A
should be connected to connector JBILA while scale B should be
connected to JBILB.
WARNING
The cable screen must always be connected to the metal
protection cap of the 9-pin connector. Do not route the
connection cable to the weighing platform together with
cables routed to power drives.