Bien Air TD 783 Instruction Download Page 1

REF 2100038-ENG/12.10

1/2

REF 2100038-ENG/12.10

TD 783 / TDS 890

STS Trimmer

S001

Without

module

Pressure of the

Pressure of the

(on the unit)

pressure gauge

pressure gauge

in bar

in bar

in bar

TD 783 / TD 783

2-hole

3

2,5

TDS 890

4-hole

4,8

2,5

4-hole REF 1600243-001

3-hole REF 1600242-001

Do not sterilise

f

ig. 1

fig. 6

fig. 4

fig. 2

fig. 7

fig. 5

fig. 3

TD 783

1600161-001

1/8“ (3.17 mm)

TD 783

1600162-001

2V

TDS 890

1600163-001

4V

REF 1000005-001

REF 1302401-010

TD 783
TDS 890

max. 25 mm

Typ 3 / ISO 1797
Code 4-6 / ISO 6360/1

min. 11 mm

ø 1.6 mm

REF 1302403-010

2.5 bar

fig. 9

fig. 10

Lubrimed
REF 1600037-006

1/2

fig. 8

1 sec.

Instruction

English

Devices entirely made in Switzerland by
Bien-Air Dental SA.

Type

Straight laboratory turbine with or without spray.

Intended use

Product intended for professional use only. Use in
for laboratory work. Any use other than that for
which this device is intended is prohibited and may
prove dangerous.

Technical data

Coupling

2- or 4 hole pressure gauge and direct

fig. 1

.

Recommanded rotation speed

Up to 300,000 rpm.

Weight

60 g (2,11 oz).

Air consumption

35 normliter/min on average.

Air spray rate

2 normliter/min to 200 kPa (29 psi).

Water spray rate

100 ml/min to 200 kPa (29 psi).

Information

The technical specifications, illustrations and di-
mensions contained in these instructions are given
only as a guide. They may not be the subject of any
claim. The manufacturer reserves the right to make
technical improvements to its equipment, without
amending these instructions. For all additional in-
formation, please contact Bien-Air Dental SA at the
address indicated on the back cover.

Using

Air pressure output
Do not use the turbine at above the required
air pressure

For use without unit:

Adjust the required air pressure to 2.5 bar (36.28
psi) with the assistance of a Bien-Air Dental inspec-
tion pressure gauge - 3-hole or 4-hole version
placed between hose and connection,

fig. 1

.

For use with the S001 unit:

Please refer to the indications of the unit’s manome-
ter. The required air pressure is 3 bar (44 psi).

Important

The device must not be started without a bur in-
serted into the chuck.

Bur

Diameter of shaft 1.60 mm (0.06 in), type 3 in ac-
cordance with ISO 1797; max. length 25 mm (0.98
in), code 4-6 in accordance with ISO 6360-1

fig. 2

.

Inserting the tool

Place the bur in the turbine’s chuck

fig. 3

. Then in-

sert the turbine head in the bur securing tool, press
the bur against the metal disc in the bur securing
tool and press firmly until you hear a «clic»

fig. 4

.

Extracting the tool

Hold the central part of the straight turbine’s han-
dle firmly in one hand and press the rear part of the
handle forward. Release and remove the bur

fig. 5

.

Seals replacement
2V:

Remove the seal according to

fig. 6

4V:

Remove the seal according to

fig. 7

Removing seal, place it on the tubes and remount
the attachment

fig. 1

Maintenance

Non-sterilisable

Only use maintenance products and components
from Bien-Air Dental. The use of other products and
components can void the guarantee.

Servicing

Never disassemble the device. For any modification
and repair, we recommend that you contact your
regular supplier or Bien-Air Dental directly.
Bien-Air Dental asks the user to have its dynamic
instruments checked or inspected at least once a
year.

Cleaning

Use

Spraynet

cleaner

fig. 8

. Spray the inside and

outside of the equipment to remove the residues and
carefully clean the surfaces.

Never immerse in disinfectant solutions.

Not suitable for ultrasonic bath.

Lubrication

Lubrication at least twice a day

.

Only use

Lubrimed

medical grease and greaser.

• First-time use: Remove the cap, screw the

knurled back-housing until the grease appears
at the end of the rod. Wipe off surplus grease
and replace cap after use.

• Remove the bur and introduce the rod of the

greaser to the bottom. The dose corresponds to
one half-turn of the knurled backhousing. Use
the guidemarks

fig. 9

.

• Replace a bur, turn on the air and run the tur-

bine for about 15 seconds.

Important

After cleaning and disinfecting the instrument
and before using it, operate at moderate
speed, with a bur in the chuck, for 10 to 15
seconds to distribute, and remove, the excess
lubricant.

Cleaning of spray tubes fig. 10

Wire cleaner.

Transport and storage conditions

Temperature between -40°C (-40°F) and 70°C
(158°F), relative humidity between 10% and 100%,
atmospheric pressure 50 kPa to 106 kPa (7.3 to 15.3
psi).

Other precautions for use

The device must be used by a qualified person in
accordance with the current legal provisions con-
cerning industrial safety, health and accident pre-

vention measures, and these working instructions.
In accordance with these requirements, the opera-
tors:
• must only use operating devices that are in per-

fect working order; in the event of irregular
functioning, excessive vibration, abnormal
heating or other signs indicating malfunction
of the device, the work must be stopped imme-
diately; in this case, contact a repair centre that
is approved by Bien-Air Dental;

• must ensure that the device is used only for the

purpose for which it is intended, must protect
themselves, their patients and third parties from
any danger, and must avoid contamination
through the use of the product.

Rest the device on a suitable support in order to
eliminate the risk of injury to you yourself and to
third parties.

Recommendations
It is essential to use dry, purified compressed
air in order to ensure the long working life of
the device. Maintain the quality of the air and
the water by regular maintenance of the com-
pressor and the filtration systems.
The use of unfiltered hard water will lead to
early blockage of the tubes, connectors and
spray cones.

After a period of prolonged use, the device must be
cleaned and serviced according to these instruc-
tions.