Bibby Sterilin Stuart SI500 Instructions For Use Manual Download Page 4

Caution: Fuses are fitted in both live and
neutral lines.

Fuses:

230V - 3.15A anti surge fuse. For all other
voltages always replace with an equivalent fuse
type and rating as those being removed,
alternatively see rating plate for fuse type and
rating.

This unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One lead has a U.K. 3 pin plug and the other
has a 2-pin “Shuko” plug for connection to the
mains. Choose the lead appropriate for your
electrical installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.

IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN

NOTE: Refer to the equipment’s rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated. The wires in the
mains cable are coloured as follows:

LIVE - 

BROWN

NEUTRAL - BLUE
EARTH - 

GREEN/YELLOW

The appropriate mains lead should be connected
to the instrument BEFORE connection to the
mains supply. Should the mains lead need
replacement a cable of 1mm

2

of harmonised

code H05VV-F connected to an IEC320 plug
should be used. N.B. the U.K. mains lead is
protected by a 10A fuse mounted in the plug
top.

IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN

4.2

Connect to the electricity supply - DO NOT

SWITCH ON.

4.3

Place the unit on a firm, level, non-slip

surface ensuring that there is sufficient free
space on all sides without coming into contact
with anything during use.

2

4.4

Open the door. Pull the two black sprung

release knobs either side of the cradle outwards
towards the sides of the incubator and draw the
cradle platform out of the incubator towards you.

4.5

Load the cradle with vessels, adjusting the

rubber-covered securing bars as required. Ensure
that the load is evenly balanced and does not
exceed the maximum load stated. 

4.6

Gently push the platform back into the

incubator until it locks in place. Close the door.

N.B: Please ensure that the retractable
platform has clicked and locked in place
before attempting to operate the unit,
failure to do this can result in damage 
to the unit.

5. Operation

5.1 Setting the speed

Using the keypad at the base of the instrument,
switch the unit ON by pressing the control knob.
Press the mode button until a red dot shows in
the display below “rpm”. The display will then
show the last set speed. Adjust the speed by
turning the control knob. The minimum speed is
30rpm and the maximum, 300rpm. 

Once the correct speed is displayed press the
start/stop button to begin motion. The speed
can be adjusted without halting the unit. The
unit can be halted at any time by pressing the
start/stop button. If the door is opened while the
unit is in motion a warning alert will sound and
the motion will stop; the word “door” will also
be momentarily displayed on the screen. Close
the door to continue.

on/off

time

rpm

start/stop

mode

Time/speed LED

Red digital display

Start/stop button

Mode button

Summary of Contents for Stuart SI500

Page 1: ...t stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Shaking Incubator SI500 Instructions for use Mode d emploi Istruzioni per uso Instrucciones de funcionamiento Ge...

Page 2: ...rbital incubator SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 English Temperature control panel Speed control panel USB connector Locating holes fo...

Page 3: ...In the case of mains interruption the unit will not restart on restoration of the electricity supply In the case of mechanical interruption e g motor stall the unit will continue operating on removal...

Page 4: ...ectricity supply DO NOT SWITCH ON 4 3 Place the unit on a firm level non slip surface ensuring that there is sufficient free space on all sides without coming into contact with anything during use 2 4...

Page 5: ...The minimum set temperature is ambient 5 C and the maximum 60 C 3 5 4 USB Connectivity Your SI500 has capability to connect to a PC via USB for long term tracking of the incubator temperature whilst...

Page 6: ...installation adaptation modification fitting of non approved parts or repair by unauthorised personnel Bibby Scientific Ltd Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 8449 360234 Fax 44 0 8...

Page 7: ...stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Panneau de commande de la temp rature Panneau de commande de la vitesse Connecteur USB Orifice de positionnement...

Page 8: ...s utiliser dans une atmosph re dangereuse ou en pr sence de substances dangereuses S assurer de l quilibre de la charge Ne pas tenter de stopper le mouvement de l appareil manuellement En cas de coupu...

Page 9: ...a plaque signal tique de l appareil afin de vous assurer que la fiche et les fusibles sont adapt s la tension et la puissance mentionn es Les couleurs des fils du cordon d alimentation sont les suivan...

Page 10: ...nutes et enfin en jours et en heures jusqu un maximum de 9 jours Lorsque l cran affiche la dur e souhait e appuyez sur le bouton start stop marche arr t afin de mettre l appareil en marche Il est poss...

Page 11: ...les des pi ces d tach es fournies ou sp cifi es par Stuart ou l un de ses repr sentants doivent tre utilis es L installation de pi ces non approuv es peut affecter l efficacit des dispositifs de s cur...

Page 12: ...deur 408 mm Largeur 422 mm Hauteur maximale des r cipients 250 mm Dimensions de la plate forme 335 x 335 mm Caract ristiques techniques Temp rature De la temp rature ambiante 5 C 60 C R solution d aff...

Page 13: ...mode rpm time on off press to start temperature C on off orbital incubator SI500 Italiano Pannello di controllo della temperatura Pannello di controllo della velocit Connettore USB Fori di riferimento...

Page 14: ...rico sia bilanciato Non fermare il movimento manualmente In caso di interruzione della rete elettrica al ripristino dell alimentazione l unit non riprende a funzionare In caso di interruzione dovuta a...

Page 15: ...V F collegato a un connettore IEC320 N B il cavo di rete tipo UK protetto da un fusibile da 10 A montato sulla sommit del connettore IN CASO DI DUBBIO CONTATTARE UN ELETTRICISTA QUALIFICATO 4 2 Colleg...

Page 16: ...Regolazione della temperatura Con il pannello di controllo superiore attivare l unit di controllo della temperatura premendo la manopola di comando Tenendo premuto il pulsante press to set premere per...

Page 17: ...ncidenti uso non conforme negligenza errori di regolazione o di riparazione danni dovuti a installazioni adattamenti modifiche montaggio di componenti non approvati o riparazioni condotte da personale...

Page 18: ...tor SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Panel de control de temperatura Panel de control de velocidad Conector USB Orificios localizadores...

Page 19: ...ed el ctrica No debe utilizarse en atm sferas peligrosas ni junto con materiales peligrosos Aseg rese de que la carga est equilibrada No intente detener el movimiento de la unidad de forma manual En c...

Page 20: ...LEVE A CABO UN ELECTRICISTA CUALIFICADO NOTA Consulte la placa de caracter sticas del equipo para asegurarse de que el enchufe y el fusible sean los adecuados para la tensi n y potencia indicadas Los...

Page 21: ...peratura pulsando el bot n de control Mantenga pulsado el bot n press to set pulsar para establecer y gire el bot n hasta que se muestre la temperatura correcta Suelte el bot n press to set pulsar par...

Page 22: ...uart o su agente La instalaci n de piezas no aprobadas puede afectar al rendimiento de las funciones de seguridad dise adas en el instrumento Para obtener una lista completa de las piezas que necesita...

Page 23: ...o de la plataforma 335 x 335 mm Especificaciones t cnicas Rango de temperatura Ambiente de 5 C a 60 C Resoluci n de visualizaci n de temperatura 0 1 C Precisi n de temperatura 0 5 C Fluctuaci n de tem...

Page 24: ...on off orbital incubator SI500 start stop mode rpm time on off press to start temperature C on off orbital incubator SI500 Temperaturbedienfeld Drehzahlbedienfeld USB Schnittstelle Positionierungsloc...

Page 25: ...tromversorgung angeschlossen ist Betreiben Sie das Ger t nicht in einer gef hrlichen Umgebung oder einer Umgebung mit gef hrlichen Werkstoffen Stellen Sie eine ausgeglichene Last sicher Versuchen Sie...

Page 26: ...F NUR VON EINEM QUALIFIZIERTEN ELEKTRIKER DURCHGEF HRT WERDEN HINWEIS Pr fen Sie anhand des Ger teschildes dass Stecker und Sicherungen f r die angegebene Spannungs und Stromst rke geeignet sind Die L...

Page 27: ...en Bedienfeld dr cken Halten Sie die Taste press to set zum Einstellen dr cken gedr ckt und drehen Sie dabei den Knopf bis die richtige Temperatur angezeigt wird Lassen Sie die Taste press to set zum...

Page 28: ...ng nicht genehmigter Teile kann die Leistung der in das Ger t integrierten Sicherheitseinrichtungen beeintr chtigen Um eine umfassende Ersatzteilliste anzufordern die Servicetechniker f r interne Repa...

Page 29: ...0 mm Tischgr e 335 x 335 mm Technische Daten Temperaturbereich Umgebungstemperatur 5 C bis 60 C Aufl sung der Temperaturanzeige 0 1 C Temperaturgenauigkeit 0 5 C Temperaturschwankung 0 5 C bei 37 C Te...

Page 30: ...Notes...

Page 31: ...be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that injury damage or l...

Page 32: ...ambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketing bibby scientific it www bibby scientific it Bibby Scientific US Ltd 3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel 800 2...

Reviews: