M800 Highspeed Handpiece 
Operator’s Manual

R e d : C

M

Y

K

B l u e : C

M

Y

K

: 0 ,

: 8 7 ,

: 9 9 ,

: 0

: 7 0 , : 7 , : 5 , : 0

The Next Standard.

Table of Contents

•  Before You Begin

•  Safety Information

•  Technical Specifications 

•  Handpiece Installation

•  Friction Grip Bur Specifications

Before You Begin 

Please  read  this  Operator’s  Manual  prior  to  using  your  AirLight  Highspeed 

Handpiece.  The use of this product is intended for use by trained dental personnel. 

We recommend that you retain this manual for future reference.  

Safety Information

•  Highspeed  handpieces  should  be  sterilized  after  each  patient  to  prevent  any  

transmission of infective organisms.

•  Suitable hygienic measures must be taken to prevent cross contamination between 

patients, operators and other personnel.

•  Only  operate  this  highspeed  handpiece  when  the  bur  is  properly  inserted  in  the 

chuck. Do not attempt to extend the length of the bur by improper positioning in 

the chuck.

•  Bur Insertion & Removal

•  Cleaning & Lubrication 

•  Troubleshooting
•  Warranty 

Handpiece Installation  (fig.1)

Note:This  handpiece  should  only  be  used  on  dental  delivery  systems  that  provide 

filtered  and  regulated  air  and  water.  First  ensure  EZ-Swivel  (HP3018)  coupling  is 

firmly attached to the handpiece tubing. 

Connection (fig. 1)

To Connect the handipece to the coupling align the handpiece to the coupling and 

push together.

Disconnection(fig.1)

Release the handpiece from the coupling and remove the handpiece.

Water & Air Adjustment

Coolant Water: Adjust the water volume at the delivery unit to obtain a satisfactory 

spray mist.

Air  Pressure:  The  operating  pressure  is  35  psi.  It  is  highly  recommended  that  the 

operating  pressure  initially  be  set  using  a  handpiece  pressure  gauge  due  to  the 

pressure loss caused by the handpiece tubing material and length.

Warning: Do not exceed recommended operating pressure as this will shorten turbine 

life and may cause additional damage to the handpiece.

Handpiece Operation

Use  standard  length  concentric  diamond  and  carbide  burs  with  .0628”  nominal 

diameter friction grip shanks designed for high speed handpieces.

Use  only  burs  with  hardened  tempered  steel  shanks.  Before  use,  verify  concentric 

rotation by operating handpiece outside the oral cavity in a safe place.

Warning: Do not use burs that produce excessive noise or do not run concentrically.  

Do not use burs that are modified, eccentrically designed, or have bent or ‘soft’ (non-

heat treated) shanks.

Friction Grip Bur Specifications  (fig.2)

For application details and operating data of the different rotary instruments please refer 

to  the  information  provided  by  the  manufacturer.The  pushbutton  autochuck  turbines 

have the following bur requirements:

Use only instruments whose shanks comply with the requirements of ISO 1797-1.

•  Shank diameter of 1.59—1.60mm

•  Overall length up to max. 25mm

•  Overall length up to max. 21mm (mini)

•  Working diameter up to max. 2.1mm (ISO 021)

CAUTION! Instruments that differ from the specified data must not be used.  Otherwise the 

mechanism of the clamping system could be damaged.

fig.1 AirLight Handpiece Installation

fig.2 Bur Specifications

   

Head Type

   

Head Size WidthxHeight

   

Optics

   

Burs

   

Chuck Type

   

Drive Air Pressure

  

 

Rotation Speed

   

Friction Grip Short Shank

   ISO 1791-1 Type3 1.59~1.60

   

Direct-LED

   

Push Button

   

Minimum 35 psi

   

340,000-370,000RPM

   

Standard

   

Mini

   

Push Button

   

Minimum 35 psi

   

380,000-410,000RPM

   

Friction Grip Short Shank

   ISO 1791-1 Type3 1.59~1.60

   

Direct-LED

   

M800-S

   

M800-M

a

b

Tubing

Pull back the connecting ring

   

12.2mmx10.8mm

   

20.6mmx13.1mm

AirLight

EZ-Swivel (HP3018)

Technical Specifications

WARNING!

(Risk of injury)

!

!

!

!

!

!

!

!

ATTENTION!

(To prevent damage)

Thermal washer

disinfectable

Sterilizable

up to the stated

temperature

135

o

C

mm