M600 & M700 AirLight Highspeed Handpiece Series

Operator’s Manual

The next standard.

Table of Contents

• 

Before You Begin

• 

Safety Information

• 

Technical Specification 

• 

Handpiece Installation

• 

Friction Grip Bur Specifications

Before You Begin 

Please read this Operator’s Manual priorto using yourAirLightHighspeedHandpiece.  
Theuse of this product is intended for use by trained dental personnel. We recom

-

mend that you retain this manual for future reference.  

Safety Information

• 

Highspeedhandpieces should be sterilized after each patient to prevent any 

transmission of infective organisms such as HIV or Hepatitis B.
• 

Suitable  hygienic  measures  must  be  taken  to  prevent  cross  contamination 

between patients, operators and other personnel.
• 

Only  operate  this  highspeedhandpiece  when  the  bur  is  properly  inserted  in 

the chuck. Do not attempt to extend the length of the bur by improper positioning in 
the chuck.

• 

Bur Insertion & Removal

• 

Cleaning & Lubrication 

• 

Troubleshooting

• 

Warranty 

Handpiece Installation 

 (

fig.1)

Note:

This handpiece should only be used on dental delivery systems that provide fil

-

tered and regulated air and water. Filtration of 25 microns is recommended.

Water & Air Adjustment

Coolant  Water  :Adjust  the  water  volume  at  the  delivery  unit  to  obtain  a  satisfactory 
spray mist.

Air Pressure :

  The  operating  pressure  is  35  psi.  It  is  highly  recommended  that  the 

operating pressure initially be set using a handpiece pressure gauge due to the pres

-

sure loss caused by the handpiece tubing material and length.

Warning:

 Do not exceed recommended operating pressure as this will shorten turbine 

life and may cause additional damage to the handpiece.

Handpiece Operation

Use standard length concentric diamond and carbide burs with .0628” nominal diam

-

eter friction grip shanks designed for high speed handpieces.

Use  only  burs  with  hardened  tempered  steel  shanks.  Before  use,  verify  concentric 
rotation by operating handpiece outside the oral cavity in a safe place.

Warning:

 Do not use burs that produce excessive noise or do not run concentrically.  

Do not use burs that are modified, eccentrically designed, or have bent or ‘soft’ (non-
heat treated) shanks.

Friction Grip Bur Specifications  

(fig.2)

For application details and operating data of the different rotary instruments please refer 
to the information provided by the manufacturer.The pushbutton autochuck turbines have 
the following bur requirements:
Use only instruments whose shanks comply with the requirements of ISO 1797-1.

• 

Shank diameter of 1.59—1.60mm

• 

Overall length up to max. 25mm

• 

Overall length up to max. 21mm (mini)

• 

Working diameter up to max. 2.1mm (ISO 021)

CAUTION! 

Instruments that differ from the specified data mustnot be used.  Otherwise the 

mechanism of the clamping system could be damaged.

fig.1 

AirLight Handpiece Installation

fig.2 

Bur Specifications