background image

22

Sling Model

Sling Size

MAA4000M-M

M

MAA4000M-L

L

MAA4000M-XL

XL

Sling Model

Sling Size

MFA4000M-M

M

MFA4000M-L

L

MFA4000M-XL

XL

Cradle Clip Sling

Cradle Clip Disposable Sling

21

If the patient is pregnant or has skin conditions, consult with 

a doctor before use.

ALWAYS check that sling weight capacity is suitable for the 

patient

All slings are type BF applied parts

Care Transfer Sling

Summary of Contents for PL350CT

Page 1: ...I E N T L I F T M o d e l P L 3 5 0 C T Ensure the product has been assembled according to the instructions in this manual All operators should read and understand the instructions for safe and proper...

Page 2: ...___ Product Serial Number _____________________________________ Date of Purchase _________________________________________ This product has been tested by SGS an accredited testing institute to comply...

Page 3: ...tional battery pack you will never have to be worried about your lift not being charged and ready for use The PL350CT is intended to be used with patients who are completely immobile or totally depend...

Page 4: ...physical contact with the patient in order for the device to carry out its intended Do Not Iron Item may not be smooted or finished with an iron Class II symbol To identify equipment is designed that...

Page 5: ...wise severely handicapped The patient lift should be used solely for transferring a user patient from one utility beds chairs toilets etc to another The patient lift should not be used for transportin...

Page 6: ...nt DOES NOT consist of any materials or ingredients to which patients are exposed that may cause bodily harm or injury Prolonged exposure or physical contact with any part of the equipment WILL NOT ca...

Page 7: ...D IN THIS DRAWING IS THE SOLE PROPERTY OF BESTCARE LLC ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF BESTCARE LLC IS PROHIBITED 9 8 2017 1 of 1 DaTong Wrench ALL DIMENSIONS...

Page 8: ...1729 2 68 1 B 1070 8 42 2 C 658 4 25 9 D 492 2 19 4 E 823 2 32 4 F 766 9 30 2 G 576 7 22 7 H 333 3 13 1 I 496 9 19 6 J 275 10 8 K 442 7 17 4 L 118 5 4 7 M 20 3 0 8 PL350CT A B C 600 D E 700 G H F M L...

Page 9: ...1608 mm H Turning Diameter 55 7 1414 mm Specifications of Control Unit Input Voltage 220 240VAC 50Hz 110 120VAC 60Hz Output Voltage 24 VDC Battery Pack 24 VDC 5AH Lift Control Handset and Panel Prote...

Page 10: ...he mast Make sure the tab on the stand bracket insert into top hole on the mast and insert the screw to secure the stand bracket as follows Step 3 Remove the quick release pin that is securing the mas...

Page 11: ...the mast Step 6 Remove the pin that is securing the boom Step 7 Remove lower actuator pin on the mast Step 8 Move the boom out of the way and install the lower end of the actuator to the mast using a...

Page 12: ...10 Insert the spreader bar into the tube on the boom as follows Align the holes and secure the spreader bar using a pin and retainer ring WARNING The tube on the boom and the spreader bar are intended...

Page 13: ...r Control Unit Installation 4 3 2 1 5 M5 SCREWS M4 SCREWS UNLESS OTHERWISE SPECIFIED PART NUMBER DATE DRAWN BY 2920 PACIFIC DRIVE NORCROSS GA 30071 9 27 2017 ALL DIMENSIONS IN MM 1 6 6 30 30 120 0 1 0...

Page 14: ...p The Lift For Storage or Transport Step 1 Remove The Spreader Bar WARNING actuator should be fully retracted boom in lowest position prior to getting ready for storage Step 2 Detach The Actuator 13 1...

Page 15: ...Step 3 Fold Up The Boom Step 4 Fold Up The Mast 14 1 2 3 1 2 3...

Page 16: ...Standing Up Lying Down 15 Warning The lift may fall and cause bodily injury if it is not secured against the wall...

Page 17: ...be one that serves the needs of the patient while providing the patient with optimal safety Never interfere with the lift unless instructed by the attendant Have a doctor nurse or medical attendant ex...

Page 18: ...ton LCD Display Panel Battery Pack Charger DC In Not Used Hand Control Charging Cradle Optional Additional Battery Pack Optional WARNING Contains a long charger cord that may cause strangulation if us...

Page 19: ...gns below The DC Charger must be connected The emergency stop button must be released The battery will not be charged if the emergency button is pressed in LCD Display Panel Signs When the emergency s...

Page 20: ...19 LED Indicator Green In Use Blank Standby Lifting Lowering Operating Lift Using Hand Control Operating Lift Using LCD Display Panel...

Page 21: ...w the LCD Display Panel will show the sign on the left either when the emergency stop button is released or when a button on the LCD Display Panel is pressed Lift operation is deactivated Battery need...

Page 22: ...0M M M MFA4000M L L MFA4000M XL XL Cradle Clip Sling Cradle Clip Disposable Sling 21 If the patient is pregnant or has skin conditions consult with a doctor before use ALWAYS check that sling weight c...

Page 23: ...ation and support please contact your dealer or visit our website at www bestcarellc com Slide the clip over the sling hook on the spreader bar through the hole Pull down the clip to lock the sling on...

Page 24: ...the patient onto the surface 11 Reverse the above procedures when returning the user patient to bed Transfer From Wheelchair to Bed 1 Help the user lean forward slightly then slide the sling down betw...

Page 25: ...tions are positioned midway under the thighs to provide good support and greater comfort 8 Position the boom to the proper height to extend the spreader bar into the car without hurting the patient or...

Page 26: ...As Needed Boom Spreader Bar Mast Base Foot Pedal Actuator Control Box Sling Sling Hardware The operator of the lift shall inspect the BestLift before each use Check all bolts for tightness Make sure...

Page 27: ...previous use Use isopropyl alcohol for any disinfection purpose Cleaning and Disinfecting Cleaning of the lift Cleaning of the sling Solvents or strong liquids or abrasive cleaning products will dama...

Page 28: ...g on usage The actuator will either go up or down but not both Replace your hand control as it is at the end of its life cycle Your control box may be malfunctioning There is a grinding sound inside t...

Page 29: ...cy Lowering Device is located at the top of the actuator shaft It is intended for use if the actuator or electronics fail to operate while the patient is suspended in midair The device consists of a p...

Page 30: ...e date for the specified time period WARRANTY PERIOD PRODUCT OR COMPONENT Three 3 Years Patient lift frame or spreader bar Two 2 Years Actuator control box pendant charger weigh scale Stand Aids all c...

Page 31: ...NOTES 30...

Page 32: ...Bestcare LLC 2920 Pacific Drive Norcross GA 30071 Ver 2020 02 www bestcarellc com 2019 Bestcare LLC...

Reviews: