Chapter 5. Configuration
46
MicroPolar (++) LB 566
5.3
Configure Plausibility
Starting from the main menu, go to the display shown on the
left via:
| SETUP | CONFIGURATION | PLAUSIBILITY | PHASE MEAS-
URE |
The display on the left shows the default settings, which may
have to be entered.
During normal measurement operation,
Sigma
lies between
0 to 100. Therefore, Sigma
max
= 100 is a good choice for
most applications. However, if you constantly get sigma val-
ues up to 300 during the measurement, you may increase
the limit up to 500. Higher sigma values than that usually in-
dicate a fault, such as continuous air bubbles which have to
be eliminated.
For most applications, the
Phi/att ratio
is not known; there-
fore, it has to be determined once. For this, you need the Au-
to setting, which otherwise must always be set to OFF.
The exact procedure is described in
chapter 9.1.
1 | - | Phase measure | 07.05 – 13:25
Sigma max
100.00
Phi/att ratio
6.00
Auto set
OFF
⌂◄
▲
▼
►
Summary of Contents for MicroPolar LB 566
Page 2: ......
Page 3: ......
Page 16: ......
Page 34: ......
Page 50: ......
Page 64: ...Chapter 7 Certificates 64 MicroPolar LB 566 ...
Page 65: ...Chapter 7 Certificates MicroPolar LB 566 65 7 2 Frequency License ...
Page 66: ...Chapter 7 Certificates 66 MicroPolar LB 566 ...
Page 67: ...Chapter 7 Certificates MicroPolar LB 566 67 ...
Page 68: ...Chapter 7 Certificates 68 MicroPolar LB 566 ...
Page 69: ...Chapter 7 Certificates MicroPolar LB 566 69 ...
Page 70: ...Chapter 7 Certificates 70 MicroPolar LB 566 ...
Page 71: ...Chapter 7 Certificates MicroPolar LB 566 71 ...
Page 72: ...Chapter 7 Certificates 72 MicroPolar LB 566 ...
Page 74: ...Chapter 8 Technical Drawings 74 MicroPolar LB 566 ...
Page 79: ...Chapter 8 Technical Drawings MicroPolar LB 566 79 8 4 2 Type LB 5650 02 ...
Page 80: ...Chapter 8 Technical Drawings 80 MicroPolar LB 566 8 4 3 Type LB 5650 03 ...
Page 81: ...Chapter 8 Technical Drawings MicroPolar LB 566 81 8 4 4 Type LB 5650 04 ...
Page 82: ...Chapter 8 Technical Drawings 82 MicroPolar LB 566 8 4 5 Type LB 5650 05 ...
Page 83: ...Chapter 8 Technical Drawings MicroPolar LB 566 83 8 4 6 Type LB 5650 09 ...
Page 84: ...Chapter 8 Technical Drawings 84 MicroPolar LB 566 8 4 7 Installation Situation in Pipelines ...
Page 86: ...Chapter 8 Technical Drawings 86 MicroPolar LB 566 8 5 2 Type LB 5651 02 ...
Page 87: ...Chapter 8 Technical Drawings MicroPolar LB 566 87 8 5 3 Type LB 5651 03 ...
Page 88: ...Chapter 8 Technical Drawings 88 MicroPolar LB 566 8 5 4 Installation Situation in Pipelines ...
Page 91: ...Notes MicroPolar LB 566 91 Notes ...
Page 92: ...Notes 92 MicroPolar LB 566 Notes ...
Page 94: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 132: ......
Page 146: ......
Page 164: ......
Page 168: ......
Page 176: ......